The Complete Radiation Dose Calculator
Every modality. Every patient. Every number cited. Effective, organ, fetal and skin dose for Nuclear Medicine, X-Ray, Fluoroscopy and CT.
For abdomen and pelvis CT scans only.
When to use this: the scan did not fully cover the abdomen and pelvis, or the fetus was near or outside the edge of the scanned range. Hardy 2019 assumes a full abdomen and pelvis scan; Damilakis et al. 2010 handles any scan range.
Needs: CTDI free in air measured at the patient’s kVp and mAs, tube voltage and beam collimation, pitch, where the scan started and ended relative to the conceptus, maternal perimeter, and (early pregnancy only) conceptus depth.
Limits: coefficients were simulated on Siemens Sensation 16 and 64 scanners with fixed tube current and mathematical phantoms; TLD checks agreed within 19%. For other scanners the authors point to a CTDIw to CTDI free in air correction published separately (Damilakis et al., Radiology 2010;257:483).
Choose two to four exams, then press Compare. Typical published adult values; Nuclear Medicine uses the activity you enter.
Angled (oblique) runs: how to enter them
- Mild angles: keep them with the PA runs on the same skin site. Jones and Pasciak recommend one skin site for modest C-arm rotation.
- Clearly separate angles: enter them as their own part (Multiple Parts). Choose a new skin site only if the fields clearly do not overlap; if unsure, assume they overlap (NCRP Report 168).
- Source to skin: measure along the angled beam to where it enters the skin.
- Table and pad: keep the transmission factor if the beam passes through the table; use 1 if it misses the table.
Where to get the source to skin distance
Tube under the table: the skin rests on the pad, so use the source to tabletop distance (the table height). Get it from the dose report or RDSR, the DICOM “Distance Source to Patient” tag (vendor dependent; on the Siemens units in Jones et al. 2014 it was table height plus a 15 cm default and had to be corrected), or have the technologist measure it once a set Ka,r is reached (Jones and Pasciak 2012; NCRP Report 168).
Lateral: the skin is about half the patient’s width from the table center, so patient size matters here. Use the source to isocenter distance minus half the patient’s lateral width if the patient was centered.
For angled runs, measure along the beam to where it enters the skin.
What the Jones 2014 defaults are
| Backscatter factor For embolization on modern C-arms, backscatter varies within a narrow range, about 1.3 to 1.5, so one value is reasonable. | 1.4 |
| f factor (tissue to air) One value for all procedures. | 1.06 |
How accurate: across 41 procedures checked against film, the simplified method with a single backscatter factor averaged 12.6% from measured peak skin dose; the full method averaged 11.8%.
When to choose another option: if you know the beam quality and DAP, use From Beam Quality; if you have measured values for your system, use Enter My Own.
Added
Breakdown by Procedure
Show the Math
Organ Doses
About These Estimates and Sources
These are reference phantom, population average estimates, not patient specific doses. Effective dose is a radiation protection quantity for comparing exposures, not for predicting an individual’s risk, and it is not meaningful for radionuclide therapy. Embryo/fetal doses carry large uncertainties.
Nuclear Medicine
- Stabin MG, Siegel JA. RADAR dose estimate report: a compendium of radiopharmaceutical dose estimates based on OLINDA/EXM version 2.0. J Nucl Med 2018;59:154–160 (electronic compendium, April 23, 2018).
- Stabin MG. New-generation fetal dose estimates for radiopharmaceuticals. J Nucl Med 2018;59:1005–1006, Supplemental Table 1.
- U.S. FDA prescribing information (DailyMed), Radiation Dosimetry sections: Amyvid, Axumin, Cerianna, Detectnet, Illuccix, Lutathera, Lymphoseek, Netspot, Neuraceq, Pluvicto, Posluma, Pylarify, Tauvid, Vizamyl, Xofigo.
X-Ray, Fluoroscopy and CT – Effective Dose
- Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M. Effective doses in radiology and diagnostic Nuclear Medicine: a catalog. Radiology 2008;248:254–263.
- Wall BF, Haylock R, Jansen JTM, et al. Radiation risks from medical X-Ray examinations as a function of the age and sex of the patient. HPA-CRCE-028. Health Protection Agency; 2011.
- Mahesh M, Ansari AJ, Mettler FA Jr. Patient exposure from radiologic and Nuclear Medicine procedures in the United States and worldwide: 2009–2018. Radiology 2023;307(1):e221263.
- Deak PD, Smal Y, Kalender WA. Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology 2010;257:158–166.
- AAPM Report No. 96. The measurement, reporting, and management of radiation dose in CT. 2008.
Embryo/Fetal Dose – X-Ray and CT
- McCollough CH, Schueler BA, Atwell TD, et al. Radiation exposure and pregnancy: when should we be concerned? RadioGraphics 2007;27:909–918.
- HPA, RCR, SCoR. Protection of pregnant patients during diagnostic medical exposures to ionising radiation. 2009.
- Hardy AJ, Angel E, Bostani M, Cagnon C, McNitt-Gray M. Estimating fetal dose from tube current-modulated (TCM) and fixed tube current (FTC) abdominal/pelvis CT examinations. Med Phys 2019;46:2729–2743.
- AAPM Report No. 204. Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. 2011.
- Damilakis J, Tzedakis A, Perisinakis K, Papadakis AE. A method of estimating conceptus doses resulting from multidetector CT examinations during all stages of gestation. Med Phys 2010;37:6411–6420.
CT Size Specific Dose Estimate
- AAPM Report No. 220. Use of water equivalent diameter for calculating patient size and size-specific dose estimates (SSDE) in CT. 2014.
- AAPM Report No. 293. Size-specific dose estimate (SSDE) for head CT. 2019.
- AAPM Report No. 246. Estimating patient organ dose with computed tomography: a review of present methodology and required DICOM information. 2019.
Fluoroscopy
- Jones AK, Pasciak AS. Calculating the peak skin dose resulting from fluoroscopically guided interventions. Part I: Methods. J Appl Clin Med Phys 2011;12(4):231–244; Erratum J Appl Clin Med Phys 2014;15(4):402.
- Jones AK, Pasciak AS. Calculating the peak skin dose resulting from fluoroscopically guided interventions. Part II: Case studies. J Appl Clin Med Phys 2012;13(1):174–186.
- Jones AK, Ensor JE, Pasciak AS. How accurately can the peak skin dose in fluoroscopy be determined using indirect dose metrics? Med Phys 2014;41(7):071913.
- International Atomic Energy Agency. Dosimetry in diagnostic radiology: an international code of practice. Technical Reports Series No. 457. Vienna: IAEA; 2007 (Appendix VIII).
- National Council on Radiation Protection and Measurements. NCRP Report No. 168: Radiation dose management for fluoroscopically-guided interventional medical procedures. 2010.
- U.S. Code of Federal Regulations, 21 CFR 1020.32, Fluoroscopic equipment (current as of October 2026).
Each result row is labeled with its effective dose basis. Most values use ICRP Publication 103 tissue weighting factors; Mettler et al. 2008 values and AAPM Report 96 combined region CT values use ICRP Publication 60. The two differ modestly for most exams, so totals that combine them are approximate.
Interpretation
- American College of Obstetricians and Gynecologists. Committee Opinion No. 723: Guidelines for diagnostic imaging during pregnancy and lactation. Obstet Gynecol 2017;130:e210–e216 (reaffirmed 2026).
- National Council on Radiation Protection and Measurements. NCRP Report No. 160: Ionizing radiation exposure of the population of the United States. 2009.
Privacy: no information is collected or stored. This calculator runs entirely in your web browser. Nothing you enter is transmitted to, collected by, or stored by Olympic Health Physics, Inc. or this website, and it is cleared when you reload or close the page.
Disclaimer and limitation of liability. This calculator is provided by Olympic Health Physics, Inc. for general informational and educational purposes only, “as is” and without warranty of any kind, express or implied, including any warranty of accuracy, completeness, or fitness for a particular purpose. It is not professional, medical, legal, or regulatory advice and does not replace the judgment of your Authorized User, Radiation Safety Officer, or qualified medical physicist. You are solely responsible for verifying all inputs, assumptions, and results; for confirming that the method used is appropriate for your situation; and for complying with applicable regulations, your license conditions, and any Agreement State requirements. Use of this calculator is entirely at your own risk. To the fullest extent permitted by law, Olympic Health Physics, Inc. and its officers, employees, and agents are not liable for any direct, indirect, incidental, consequential, or other damages arising from the use of, or reliance on, this calculator or its results. Using this calculator means you accept these terms.
Version 2 · October 18, 2026